Categories
Categories
My Cart
We use cookies in line with legal regulations to improve your shopping experience. For details, see our Cookie Policy.
Testo 176 T3 4-Channel Thermocouple Temperature Data Logger – Type K/T/J, −200…+1000 °C, Metal Housing, IP65, Probe and USB Cable Not Included (0572 1763)
Testo 176 T3 4-Channel Thermocouple Temperature Data Logger – Type K/T/J, −200…+1000 °C, Metal Housing, IP65, Probe and USB Cable Not Included (0572 1763)
Testo 176 T3 4-Channel Thermocouple Temperature Data Logger – Type K/T/J, −200…+1000 °C, Metal Housing, IP65, Probe and USB Cable Not Included (0572 1763)
Testo 176 T3 4-Channel Thermocouple Temperature Data Logger – Type K/T/J, −200…+1000 °C, Metal Housing, IP65, Probe and USB Cable Not Included (0572 1763)
Testo 176 T3 4-Channel Thermocouple Temperature Data Logger – Type K/T/J, −200…+1000 °C, Metal Housing, IP65, Probe and USB Cable Not Included (0572 1763)
Testo 176 T3 4-Channel Thermocouple Temperature Data Logger – Type K/T/J, −200…+1000 °C, Metal Housing, IP65, Probe and USB Cable Not Included (0572 1763)
Testo 176 T3 4-Channel Thermocouple Temperature Data Logger – Type K/T/J, −200…+1000 °C, Metal Housing, IP65, Probe and USB Cable Not Included (0572 1763)

Testo 176 T3 4-Channel Thermocouple Temperature Data Logger – Type K/T/J, −200…+1000 °C, Metal Housing, IP65, Probe and USB Cable Not Included (0572 1763)

  • Stock: In Stock
  • Brand: Testo
  • Model: 0572 1763
$696.63 (VAT incl.)
Ex Tax: $580.52
Fast Dispatch
Shipped from Türkiye by international express courier (see delivery page)
Secure Shopping
Your card details cannot be seen by anyone, including our staff.
Shipping
Your discounted shipping rate is calculated in the cart.

Testo 176 T3 (product code 0572 1763) is a multi-channel temperature data logger with a metal housing that simultaneously measures and records temperature at four different points using type K, T and J probes connected to 4 external thermocouple inputs. Depending on the connected probe type, it measures in the range of −200 … +1000 °C with type K, −200 … +400 °C with type T and −100 … +750 °C with type J, with 0.1 °C resolution, and stores up to 2 million readings. With its rugged IP65 protected housing, it is used to monitor compliance with specified temperature values in industrial processes.

Important: The device requires thermocouple probes for measurement; the thermocouple probes and the mini USB connection cable required to program the logger are not included in the scope of delivery. Type K, T or J probes suitable for the application and, if you do not have one at first purchase, a mini USB cable must be obtained separately.

Simultaneous Thermocouple Temperature Logging at Four Points

The active channels and the probe type to be connected to each input are defined with the free Testo ComSoft Basic software; the probes are plugged into the inputs numbered 1 to 4 on the underside of the housing. The measuring interval can be freely selected between 1 second and 24 hours; at a 15-minute measuring interval it runs for approximately 8 years on a single lithium AA battery, and the expected battery life is calculated in the software according to the measuring interval and the number of connected sensors. Recorded data are retained when the battery runs out or is being replaced.

The device has no display; operating status is monitored with colored LEDs: the red LED flashes when the remaining battery life falls below 30 and 10 days and when the battery is empty, and flashes three times when the button is pressed if a limit value has been exceeded; the yellow LED indicates logging mode and the green LED standby mode. Data are transferred to a computer via the mini USB port on the side or to a card inserted in the SD card slot: when an SD card is inserted and the GO button is pressed for more than 2 seconds, copying starts; the yellow LED lights during the process and the process is complete when the green LED flashes twice. In the software, data are viewed as graphs and tables and exported to Excel. The device is placed in the wall bracket and secured with the supplied lock.

Technical Specifications (Official Testo Information)

  • Brand / Model: Testo 176 T3 thermocouple temperature data logger · product code 0572 1763
  • Channels: 4 external thermocouple inputs (type K, T and J)
  • Measuring Range (by probe type): Type K (NiCr-Ni) −200 … +1000 °C · Type T (Cu-CuNi) −200 … +400 °C · Type J (Fe-CuNi) −100 … +750 °C
  • Instrument Accuracy (±1 digit): ±0.3 °C (−100 … +70 °C) · ±0.5% of reading (from +70.1 °C to the upper limit of the probe type) · ±1% of reading (−200 … −100.1 °C, type K and T)
  • Resolution: 0.1 °C
  • Logging Interval: 1 second … 24 hours (2 seconds … 24 hours in online measurement)
  • Memory: 2,000,000 readings
  • Status Indication: red / yellow / green LED (no display)
  • Interface: mini USB, SD card slot · Testo ComSoft Basic (free), ComSoft Professional (optional)
  • Power Supply: 1 × special lithium AA battery (Testo spare battery code 0515 1760) · battery life 8 years (15 min measuring interval, +25 °C)
  • Operating / Storage Temperature: −20 … +70 °C / −40 … +85 °C
  • Protection Class: IP65 (with probes connected or inputs closed with blanking plugs)
  • Standard: 2014/30/EU · 2011/65/EU
  • Housing / Dimensions / Weight: metal · 103 × 63 × 33 mm · approx. 430 g

Scope of Delivery

  • Testo 176 T3 4-channel temperature data logger (metal housing)
  • Wall bracket and lock
  • Battery
  • Test protocol
  • Thermocouple probes, mini USB connection cable and SD card are not included in the scope of delivery

Technician Tips

  • Match the probe type with the software: plug the probe of the thermocouple type you defined in the software for each channel into the corresponding numbered input on the housing; seat the plugs fully with the correct polarity and without force.
  • Close unused inputs: IP65 protection is only achieved when probes are connected to the inputs or unused inputs are closed with blanking plugs.
  • Take care with non-insulated surface probes: the maximum permissible potential difference between sensor inputs is 50 V; take this into account if you use surface probes with non-insulated thermocouples.
  • Only the probe tip at the hot spot: the logger operates at −20 … +70 °C; place only the probe tip at high-temperature points, and do not expose the probe handle and cable to more than 70 °C unless a permissible value is specified. Testo offers a heat protection case (0572 9999) as an accessory, protecting up to +200 °C.
  • Protect battery life: prolonged LED flashing, reading out with an SD card several times a day and extreme temperature fluctuations shorten battery life; plan the readout frequency as needed.

Technical document: Testo 176 T3 / T4 data logger datasheet (PDF, English)

Be the first to write a review for this product.

Write a review

Note: HTML is not translated!
Bad           Good

Loading...

Unlimited Blocks, Tabs or Accordions with any HTML content can be assigned to any individual product or to certain groups of products, like entire categories, brands, products with specific options, attributes, price range, etc. You can indicate any criteria via the advanced product assignment mechanism and only those products matching your criteria will display the modules.

Also, any module can be selectively activated per device (desktop/tablet/phone), customer login status and other criteria. Imagine the possibilities. 

Prices are shown in US dollars for reference, converted at the current exchange rate. Payment is charged in Turkish lira (TL); the amount on your card statement may differ slightly. Türkçe