Testo pH 4.01 buffer solution (product code 0554 2061) is a reference solution used for the calibration and adjustment of pH meters, supplied in a practical 250 ml dosing bottle. It is used for the pH 4 calibration point on the Testo 205, Testo 206 pH1, Testo 206 pH2 and Testo 206 pH3 pH/temperature meters. According to the bottle label, the solution is traceable to N.I.S.T. and PTB, and the pH value of lot samples is verified by DKD; Testo user manuals list it as DAkkS calibration certified.
pH 4 Point in pH Meter Calibration
On the Testo 205, 1-point calibration is carried out at pH 4, 7 or 10, and 2-point calibration at pH 4–7 or pH 7–10; on the Testo 206 series, 1-, 2- or 3-point calibration can be selected. For a pH 4–7 two-point calibration, this solution is used together with the pH 7.00 buffer solution (Testo 0554 2063). In calibration mode the probe tip is immersed in the buffer solution; when the instrument finds a stable value (less than 0.02 pH change in 20 seconds) it saves the calibration point and finally displays the slope (mV/pH) and offset (mV) values.
Technical Specifications (Official Testo Data)
- Brand / Product: Testo pH 4.01 buffer solution · product code 0554 2061
- pH Value: 4.01 (+25 °C)
- pH by Temperature (bottle label): 5 °C: 4.01 · 10 … 20 °C: 4.00 · 25 … 30 °C: 4.01 · 35 °C: 4.02 · 40 °C: 4.03 · 45 °C: 4.04 · 50 °C: 4.06
- Packaging: 250 ml dosing bottle
- Traceability: traceable to N.I.S.T. and PTB · DAkkS calibration certified (Testo accessory list)
- Storage Temperature: +5 … +35 °C (bottle label)
- Compatible Instruments: Testo 205 · Testo 206 pH1 · Testo 206 pH2 · Testo 206 pH3
Scope of Delivery
- 1 × pH 4.01 buffer solution in a 250 ml dosing bottle
Technician Tips
- Do not let the glass tip touch the bottle: during calibration, the glass part of the pH probe tip must not touch the plastic of the bottle and the instrument must not be left inside the bottle; otherwise a calibration deviation of up to ±0.4 pH may occur.
- Read the label: Testo requires that the usage instructions on the buffer solution label be followed during calibration; the expiry date and lot information are on the bottle label.
- Know the temperature effect: according to the table on the label, the solution's pH value is 4.00 at 10 … 20 °C, 4.03 at 40 °C and 4.06 at 50 °C; check the value for the temperature at which you calibrate in the table.
- Renew the solution on an error code: if Er1 (invalid slope) or Er2 (invalid offset) appears after calibration, repeat the calibration; use fresh buffer solution if necessary.
- Monitor the electrode's service life: if at the end of calibration the slope is below 50 mV/pH or the offset is above 60 mV, the pH electrode is worn out and the probe tip must be replaced.
Loading...
Unlimited Blocks, Tabs or Accordions with any HTML content can be assigned to any individual product or to certain groups of products, like entire categories, brands, products with specific options, attributes, price range, etc. You can indicate any criteria via the advanced product assignment mechanism and only those products matching your criteria will display the modules.